SERVICE AED PLUS

SERAED03

SERVICE AED PLUS

SERAED03

Quantity Per Pack:
1
SERVICE AED PLUS ¥ Check for damage that may cause the machine to malfunction.¥ Perform shock test on AED using a cardiac simulator.¥ Clear data of shock test from rescue history.¥ Check condition of electrode packaging & confirm still within date¥ Check and adjust where necessary the time & date stored on the AED.¥ Certification of compliance after service.